 Installation Reports
 Installation Reports

• Introduction
This report provides the details of the installation of the U350 for the Hitachi SU8240 in 1C08-01, Building 81, 974 East Arques Avenue, Sunnyvale, CA 94085. Applied Materials hired VEC to install the vibration isolation system and provide a report of the completed work.
• Summary Results
Below is a summary of the results:
- U350 ON PASSED the Vibration specifications
• Product Details
| Part Delivered | Serial Number | Software | 
|---|---|---|
| U350 | 240614R2-1 | - | 
• Summary of Work
We installed the system by
- Positioning the daeil under the microscope
We tested the system by
- Creating and uploading a site-specific tune
- Collecting vibration data to verify the performance
We left the system
- In position and on
• Procedures
Vibration
VEC conducted vibration measurements in six directions. Acceleration measurements were taken at a bandwidth of 200Hz with units of um (Pk-Pk). The acceleration spectra with 200Hz bandwidths were recorded.
• Setup
| Vibration | |
|---|---|
| Bandwidth | 200Hz | 
| Lines | 3200 | 
| Window | Hanning | 
| Coupling | AC SE | 
| Average Type | Stable | 
| Duration | 60 seconds | 
| Engineering Units | um (Pk-Pk) | 
• Specifications
VEC used the following specifications for the project.
Specification Summary
| Tool | Specification | 
|---|---|
| Hitachi SU8240 | Vibration All Directions VC-D | 
• Results
Summary Results
Below is a summary of the results:
- U350 ON PASSED the Vibration specifications
Detailed Results
| Type | Position | Direction | Specification | Measurement | Result | 
|---|---|---|---|---|---|
| Vibration | Floor | Vertical | VC-D | VC-C | FAIL | 
| Left-Right | VC-E | PASS | |||
| Front-Back | VC-E | PASS | |||
| U 350 | Vertical | VC-E | PASS | ||
| Left-Right | VC-F | PASS | |||
| Front-Back | VC-E | PASS | 
• Data
Figure 1: Vibration (Vertical – Floor)
 
											Figure 2: Vibration (Vertical – U 350)
 
											Figure 3: Vibration (Vertical – Transmissibility)
 
											Figure 4: Vibration (Left-Right – Floor)
 
											Figure 5: Vibration (Left-Right – U 350)
 
											Figure 6: Vibration (Left-Right – Transmissiblilty)
 
											Figure 7: Vibration (Front-Back – Floor)
 
											Figure 8: Vibration (Front-Back – U 350)
 
											Figure 9: Vibration (Front-Back – Transmissibility)
 
											