fbpx

DAEIL SYSTEMS

DAEIL SYSTEMS

ZEISS Gemini FE-SEM
DVIA-MB1000 Installation Report

Installation Reports

1. Measurement Details

  • Measurement Date
    June 2, 2017
  • Measurement Devices
    1. LAN-XI Data Acquisition Hardware
    – Brüel & Kjæ r 3050-B-040
    2. Data Analysis Software
    – Brüel & Kjæ r PULSE LAB SHOP 14
    3. Sensors
    – PCB Accelerometer
    – Model: 393B05
  • Measurement Location
    1st Floor

2. Equipment Information

  • Manufacturer
    ZEISS
  • Model
    Gemini FE-SEM
  • Floor Vibration Specification
    VC-E in Vertical and Horizontal Axis

3. Vibration Isolation System Information
Model: DVIA-MB1000

DVIA-MB Base Active Vibration Isolation Platform
Platform Dimensions 1140 x 910 x 224 mm
Load Capacity 500 - 1700 kg
Actuator Electromagnetic Actuator
Maximum Actuator Force Vertical: 40N, Horizontal: 20 N
Degrees of Freedom 6 degrees
Active Isolation Range 0.5 - 100 Hz
Vibration Isolation at 2 Hz ≥90%
Vibration Isolation at 10 Hz ≥99%
Input Voltage (V) AC100 - 240V / 50 - 60 Hz / 1A
Power Consumption (W) Maximum 110W, <50 W in normal operation
Operating Range Temperature (°C) 5 - 50 °C
Humidity (%) 20 - 90%
Required Air Pressure ≥ 0.5 MPa (5 bar)

4. Installation Photo

DVIA MB1000 ZEISS Gemini SEM

5. Summary

Vibration Criterion Curves
Frequency Range 1-10 Hz 10 - 80 Hz
Measurement Direction Z axis (Vertical) X axis (Left to Right) Y axis (Front to Back) Z axis (Vertical) X axis (Left to Right) Y axis (Front to Back)
Floor VC-D (Fail) VC-D (Fail) VC-D (Fail) VC-B (Fail) VC-B (Fail) VC-C (Pass)
On Active Vibration Isolation System VC-G (Pass) VC-G (Pass) VC-G (Pass) VC-G (Pass) VC-G (Pass) VC-G (Pass)

6. Results
Z-axis (Vertical)

DVIA MB1000 ZEISS Gemini SEM z-axis-vertical

X-axis (Left to Right)

DVIA MB1000 ZEISS Gemini SEM x-axis-left-to-right

Y-axis (Front to Back)

DVIA MB1000 ZEISS Gemini SEM y-axis-front-to-back

7. Reference
Generic Vibration Criteria

Criterion Curve Description Amplitude1) μm/s (in/s) Detail Size2) μm
Workshop (ISO) Distinctly perceptible vibration, Appropriate to workshops and non-sensitive areas. 800 (32,000) N/A
Office (ISO) Perceptible vibration. Appropriate to offices and non-sensitive areas. 400 (16,000) N/A
Residential Area (ISO) Barely perceptible vibration. Appropriate to sleep areas in most instances. Usually adequate for computer equipment, hospital recovery rooms, semiconductor probe test equipment, and microscopes less than 40x. 200 (8,000) 75
Operating Theatre (ISO) Vibration not perceptible. Suitable in most instances for surgical suites, microscopes to 100x and for other equipment of low sensitivity. 100 (4,000) 25
VC-A Adequate in most instances for optical microscopes to 400x, microbalances, optical balances, proximity and projection aligners, etc. 50 (2,000) 8
VC-B Appropriate for inspection and lithography equipment (including steppers) to 3pm line widths. 25 (1,000) 3
VC-C Appropriate standard for optical microscopes to 1000x, lithography and inspection equipment (including moderately sensitive electron microscopes) to 1μm detail size, TFT-LCD stepper/scanner processes. 12.5 (500) 1-3
VC-D Suitable in most instances for demanding equipment, including many electron microscopes (SEMs and TEMs) and E-Beam systems. 6.25 (250) 0.1-0.3
VC-E A challenging criterion to achieve. Assumed to be adequate for the most demanding of sensitive systems including long path, laser-based, small target systems, E-Beam lithography systems working at nanometer scales, and other systems requiring extraordinary dynamic stability. 3.12 (125) <0.1
VC-F Appropriate for extremely quite research spaces; generally difficult to achieve in most instances, especially cleanrooms. Not recommended for use as a design criterion, only for evaluation. 1.56(62.5) N/A
VC-G Appropriate for extremely quite research spaces; generally difficult to achieve in most instances, especially cleanrooms. Not recommended for use as a design criterion, only for evaluation. 0.78(31.3) N/A
  1. As measured in one-third octave bands of frequency over the frequency 8 to 80 Hz (VC-A and VC-B) or 1 to 80 Hz (VC-C through VC-G).
  2. The detail size refers to line width in the case of microelectronics fabrication, the particle (cell) size in the case of medical and pharmaceutical research, etc, It is not relevant to imaging associated with probe technologies, AFMs, and nanotechnology.

    The information given in this table is for guidance only. In most instances, it is recommended that the advice of someone knowledgeable about applications and vibration requirements of the equipment and processes be sought.