Thermofisher Scientific PRISMA E SEM
DVIA-MB1000 Installation Report

DVIA MB1000 FEI SEM INSPECT S50 Active Vibration Isolation ON

Case Studies

1. Measurement Details

  • Measurement Date
    February 11, 2019
  • Measurement Devices
    1. LAN-XI Data Acquisition Hardware
    – Brüel & Kjæ r 3050-A-040 (Serial Number: 3050-111438)
    2. Data Analysis Software
    – Brüel & Kjæ r PULSE LAB SHOP 22
    3. Sensors
    – PCB Accelerometer
    – Model: 393B05 (Serial Number: 48995, 40626)
  • Measurement Location
    4st Floor
  • Measurement Setup
    Bandwidth: 0 – 100 Hz
    Lines: 400
    Window: Hanning
    Averaging: Fast Fourier Transform Spectrum Averaging
    Amplitude Units: m/s2
    Spectral Unit: RMS

2. Equipment Information

  • Manufacturer
    Thermofisher Scientific
  • Model
    PRISMA E SEM
  • Floor Vibration Specification
    VC-E in Vertical and Horizontal Axis

3. Vibration Isolation System Information
Model: DVIA-MB1000

DVIA-MB Base Active Vibration Isolation Platform
Model DVIA-MB1000 DVIA-MB3000 DVIA-MB6000
Platform Dimensions (L x W x H) Custom-made
Load Capacity 500 - 1700 kg 1500 - 3500 kg 3000 - 6000 kg
Actuator Electromagnetic Actuator
Maximum Actuator Force Vertical: 40 N, Horizontal: 20 N Vertical: 80 N, Horizontal: 40 N
Degrees of Freedom 6 degrees
Active Isolation Range 0.5 - 100 Hz
Vibration Isolation at 1 Hz ≥90%
Input Voltage (V) AC100 - 240V / 50 - 60 Hz / 1A
Power Consumption (W) Maximum 110W, <50 W in normal operation
Operating Range Temperature (°C) 5 - 50 °C
Humidity (%) 20 - 90%
Required Air Pressure ≥ 0.5 MPa (≥ 5 bar)

4. Summary

Floor Vibration Specification
Frequency Range 1 - 80 Hz
Floor Vibration Specification VC-E VC-E VC-E
Measurement Direction Z axis (Vertical) X axis (Left to Right) Y axis (Front to Back)
Floor Vibration Fail Fail Fail
Vibration On Active Vibration Isolation System Pass Pass Pass

5. Results – Vibration Criterion Curves
VC Curves, Z-axis (Vertical)

DVIA MB1000 FEI SEM INSPECT S50 z-axis-vertical

AVI = Active Vibration Isolation

The measured vertical floor vibration did not meet the vibration specification VC-E.
The active vibration isolation system reduced the vertical floor vibration from VC-A to VC-E.

VC Curves, X-axis (Left to Right)

DVIA MB1000 FEI SEM INSPECT S50 x-axis-left-to-right

AVI = Active Vibration Isolation

The measured X-axis floor vibration did not meet the floor vibration specification VC-B.
The active vibration isolation system reduced the X-axis floor vibration from VC-B to VC-F.

VC Curves, Y-axis (Front to Back)

DVIA MB1000 FEI SEM INSPECT S50 y-axis-front-to-back

AVI = Active Vibration Isolation

The measured Y-axis floor vibration did not meet the floor vibration specification VC-C.
The active vibration isolation system reduced the Y-axis floor vibration from VC-C to VC-G.

6. Results – Active Isolation Off vs Active Isolation On

DVIA MB1000 FEI SEM INSPECT S50 Active Vibration Isolation OFF

Active Vibration Isolation Off

HV:15.00 kV
Mag: 50 000 x
Linewidth: 1 μm

DVIA MB1000 FEI SEM INSPECT S50 Active Vibration Isolation ON

Active Vibration Isolation On

HV:15.00 kV
Mag: 50 000 x
Linewidth: 1 μm

DVIA MB1000 FEI SEM INSPECT S50 Active Vibration Isolation OFF

Active Vibration Isolation Off

HV:15.00 kV
Mag: 50 000 x
Linewidth: 1 μm

DVIA MB1000 FEI SEM INSPECT S50 Active Vibration Isolation ON

Active Vibration Isolation On

HV:15.00 kV
Mag: 50 000 x
Linewidth: 1 μm

7. Reference
Generic Vibration Criteria

Criterion Curve Description Amplitude1) μm/s (in/s) Detail Size2) μm
Workshop (ISO) Distinctly perceptible vibration, Appropriate to workshops and non-sensitive areas. 800 (32,000) N/A
Office (ISO) Perceptible vibration. Appropriate to offices and non-sensitive areas. 400 (16,000) N/A
Residential Area (ISO) Barely perceptible vibration. Appropriate to sleep areas in most instances. Usually adequate for computer equipment, hospital recovery rooms, semiconductor probe test equipment, and microscopes less than 40x. 200 (8,000) 75
Operating Theatre (ISO) Vibration not perceptible. Suitable in most instances for surgical suites, microscopes to 100x and for other equipment of low sensitivity. 100 (4,000) 25
VC-A Adequate in most instances for optical microscopes to 400x, microbalances, optical balances, proximity and projection aligners, etc. 50 (2,000) 8
VC-B Appropriate for inspection and lithography equipment (including steppers) to 3pm line widths. 25 (1,000) 3
VC-C Appropriate standard for optical microscopes to 1000x, lithography and inspection equipment (including moderately sensitive electron microscopes) to 1μm detail size, TFT-LCD stepper/scanner processes. 12.5 (500) 1-3
VC-D Suitable in most instances for demanding equipment, including many electron microscopes (SEMs and TEMs) and E-Beam systems. 6.25 (250) 0.1-0.3
VC-E A challenging criterion to achieve. Assumed to be adequate for the most demanding of sensitive systems including long path, laser-based, small target systems, E-Beam lithography systems working at nanometer scales, and other systems requiring extraordinary dynamic stability. 3.12 (125) <0.1
VC-F Appropriate for extremely quite research spaces; generally difficult to achieve in most instances, especially cleanrooms. Not recommended for use as a design criterion, only for evaluation. 1.56(62.5) N/A
VC-G Appropriate for extremely quite research spaces; generally difficult to achieve in most instances, especially cleanrooms. Not recommended for use as a design criterion, only for evaluation. 0.78(31.3) N/A
  1. As measured in one-third octave bands of frequency over the frequency 8 to 80 Hz (VC-A and VC-B) or 1 to 80 Hz (VC-C through VC-G).
  2. The detail size refers to line width in the case of microelectronics fabrication, the particle (cell) size in the case of medical and pharmaceutical research, etc, It is not relevant to imaging associated with probe technologies, AFMs, and nanotechnology.

    The information given in this table is for guidance only. In most instances, it is recommended that the advice of someone knowledgeable about applications and vibration requirements of the equipment and processes be sought.