Installation Reports
• Introduction
This report provides the details of the installation of the U350 for the Hitachi SU8240 in 1C08-01, Building 81, 974 East Arques Avenue, Sunnyvale, CA 94085. Applied Materials hired VEC to install the vibration isolation system and provide a report of the completed work.
• Summary Results
Below is a summary of the results:
- U350 ON PASSED the Vibration specifications
• Product Details
Part Delivered | Serial Number | Software |
---|---|---|
U350 | 240614R2-1 | - |
• Summary of Work
We installed the system by
- Positioning the daeil under the microscope
We tested the system by
- Creating and uploading a site-specific tune
- Collecting vibration data to verify the performance
We left the system
- In position and on
• Procedures
Vibration
VEC conducted vibration measurements in six directions. Acceleration measurements were taken at a bandwidth of 200Hz with units of um (Pk-Pk). The acceleration spectra with 200Hz bandwidths were recorded.
• Setup
Vibration | |
---|---|
Bandwidth | 200Hz |
Lines | 3200 |
Window | Hanning |
Coupling | AC SE |
Average Type | Stable |
Duration | 60 seconds |
Engineering Units | um (Pk-Pk) |
• Specifications
VEC used the following specifications for the project.
Specification Summary
Tool | Specification |
---|---|
Hitachi SU8240 | Vibration All Directions VC-D |
• Results
Summary Results
Below is a summary of the results:
- U350 ON PASSED the Vibration specifications
Detailed Results
Type | Position | Direction | Specification | Measurement | Result |
---|---|---|---|---|---|
Vibration | Floor | Vertical | VC-D | VC-C | FAIL |
Left-Right | VC-E | PASS | |||
Front-Back | VC-E | PASS | |||
U 350 | Vertical | VC-E | PASS | ||
Left-Right | VC-F | PASS | |||
Front-Back | VC-E | PASS |
• Data
Figure 1: Vibration (Vertical – Floor)
Figure 2: Vibration (Vertical – U 350)
Figure 3: Vibration (Vertical – Transmissibility)
Figure 4: Vibration (Left-Right – Floor)
Figure 5: Vibration (Left-Right – U 350)
Figure 6: Vibration (Left-Right – Transmissiblilty)
Figure 7: Vibration (Front-Back – Floor)
Figure 8: Vibration (Front-Back – U 350)
Figure 9: Vibration (Front-Back – Transmissibility)