fbpx

DAEIL SYSTEMS

DAEIL SYSTEMS

Thermo Fisher Scientific Glacios 2 with Selectris Cryo-TEM
DVIA-MB3000 Installation Report

Installation Reports

UNIVERSITY OF NORTH DAKOTA Logo

• Executive Summary

Introduction

The purpose of this work is to install the Vibration Isolation MB 3000 system for the Thermo
Fisher Scientific Talos F200X at the University of North Dakota Grand Forks. Thermo Fisher
Scientific hired VEC to install the field cancellation system and provide a report of the completed
work.

Summary Results

Below is a summary of the results:

  • MB3000 ON PASSED the Vibration specifications

• Introduction

The purpose of this work is to install the Vibration Isolation MB 3000 system for the Thermo Fisher Scientific Talos F200X at the University of North Dakota Grand Forks.
Thermo Fisher Scientific hired VEC to install the field cancellation system and provide a report of the completed work.

• Product Details

Part Delivered Serial Number Software
MB3000 221104R3 Site-Specific Tune

• Summary of Work

We installed the system by

  • Positioning the Daeil under the microscope

 

We tested the system by

  • Creating and uploading a site specific tune
  • Collecting vibration data to verify the performance

 

We left the system

  • In position and on

• Procedures

Vibration

VEC conducted vibration measurements in six directions. Acceleration measurements were taken at a bandwidth of 200 with units of um/s (RMS).
The acceleration spectra with 200 bandwidths were recorded.

• Setup

Vibration
Bandwidth 200
Lines 1600
Window Hanning
Coupling ICP 2mA
Average Type Stable
Duration 60
Engineering Units um/s (RMS)

Vibration

VEC conducted vibration measurements in three directions. Acceleration measurements were
taken at a bandwidth of Hz with units of um/s. The acceleration spectra with Hz bandwidths were
recorded.

• Setup

Part Delivered
Bandwidth Hz
Duration seconds
Engineering Units um/s

• Specifications

VEC used the following specifications for the project.

Specification Summary

PTool Specification
Thermo Fisher Scientific Talos F200X Vibration
 VC-F
  X & Y
   1.56 um/s 1Hz - 100Hz
 VC-E
  Z
   3.12 um/s 1Hz - 100Hz

Detailed Specifications

Figure 1: Thermo Fisher Scientific Talos F200X – 1 Vibration Specification

 

5.4.1 VC Curves for System Enclosure

In general VC-F should be achieved.
System Vertical Left to right Front to back
Talos F200X,S; TEM info limit 0.12nm E F F
Talos F200X,S; STEM res. 0.16nm E F F
Talos F200C; TEM info limit 0.15nm E F F
Talos F200C; STEM res. 0.20nm E F F
Vibration criteria:
VC-E • Floor a 700 kg/m2, indication > 29 cm thick
• Rigid support structure with small spans or heavy foundation plate
• Zone VC-E separately funded, keep around by dilation free of residual building
• Technical rooms, elevators, stairways and circulation areas outside dilated zone VC-E
• At the location of the dilatation disconnect channels and pipes to prevent transmission of vibration. No vibration sources in zone VC-E
• Technical rooms on distance. isolate vibration sources outside zone VC-E
• Zone VC-E on ground floor or 1st floor
VC-F • Floor a 1 000 kg/m', indication > 58 cm thick, heavy foundation plate
• Zone VC-F separately funded, keep around by dilation free of residual building
• Technical rooms, elevators, stairways and circulation areas outside dilated zone VC-F
• At the location of the dilatation disconnect channels and pipes to prevent transmission of vibration
• No vibration sources in zone VC-F
• Technical rooms on distance, isolate vibration sources outside zone VC-F
• Zone VC-F on ground floor

• Results

Summary Results

Below is a summary of the results:

  • MB3000 PASSED the Vibration specifications
 

Detailed Results

Type Position Direction Specification Measurement Result
Vibration Floor X-Axis (VC-F) 1.56 um/s 1Hz - 100Hz 1 um/s @ 63 Hz PASS
Y-Axis (VC-F) 0.515 um/s @ 63 Hz PASS
Z-Axis (VC-E) 3.12 um/s 1Hz - 100Hz 3 um/s @ 63 Hz PASS
Isolators X-Axis (VC-F) 1.56 um/s 1Hz - 100Hz 0.08 um/s @ 63 Hz PASS
Y-Axis (VC-F) 0.107 um/s @ 3 Hz PASS
Z-Axis (VC-E) 3.12 um/s 1Hz - 100Hz 0.133 um/s @ 32 Hz PASS

• Data

Thermo Fisher Scientific Talos F200X

Vibration

Figure 2: Ground X-Axis

 
Measurement:1 um/s @ 63 Hz

Figure 3: Ground Y-Axis

Measurement:0.515 um/s @ 63 Hz

Figure 4: Ground Z-Axis

Measurement:3 um/s @ 63 Hz

Figure 5: Isolators X Axis

Measurement:0.08 um/s @ 63 Hz

Figure 6: Isolators Y Axis

Measurement:0.107 um/s @ 63 Hz

Figure 7: Isolators Z Axis

Measurement:0.133 um/s @ 63 Hz

Figure 8: Transmissibility X Axis

Figure 9: Transmissibility Y Axis

Figure 10: Transmissibility Z Axis